Delivered Energy at 200J Defibrillator Setting into a Range of Loads

Note:

For a complete listing of X Series

Delivered Energy at Every Defibrillator Setting into a Range

of Loads

, see Appendix A in the X Series Operator’s Guide.

Selected

Energy

Load

25

50

75

100

125

150

175

Accuracy*

5 J

3 J

5 J

6 J

6 J

6 J

6 J

6 J

±15%

50 J

35 J

54 J

59 J

61 J

62 J

61 J

59 J

±15%

100 J

71 J

109 J

119 J

122 J

125 J

123 J

119 J

±15%

200 J

142 J

230 J

249 J

253 J

269 J

261 J

260 J