0 nibp calibration check, Nibp calibration check – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 43
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
37
20.0 NIBP Calibration Check
Equipment
NIBP Simulator, dual lumen hose w/ NIBP test fixture cuff (9100-3072-TF) or equivalent
Note:
If ECG simulator is connected to the device, either disconnect ECG simulator or
disable the Smart Cuff feature in the Supervisor Menu.
Setup
Attached the hose connected to the T-fitting of the NIBP test fixture cuff to the pressure port of
the NIBP simulator. Set simulator to Static Pressure.
Do This...
Observe This...
Pass/Fail
20.1 Turn on the X Series unit.
Note: If the unit is in AED mode, press
the Manual quick access key to go into
manual mode.
20.2 Connect a dual lumen hose from the NIBP
simulator to the NIBP port of the X
Series.
20.3 On the X Series unit, do the following to
enter the NIBP Test control panel:
•
Press the More (
) then the Setup
(
) quick access keys.
•
Use the navigation keys to select
Supervisor>Service>Device
Tests>NIBP Test.
Note: You need a supervisor passcode to
enter the Supervisor menu.
20.4 From the X Series NIBP Test control
panel, use the navigation keys to select:
•
Close Valve
•
Turn Pump On
20.5 Monitor the pressure reported by the
external pressure standard.
20.6 Once the pressure reaches the target
pressure of 50 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.7 Allow the system to stabilize for at least
10 seconds.
20.8 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
o
o
20.9 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
Reading should be within
4mmHg.
o
o