Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

37

20.0 NIBP Calibration Check

Equipment

NIBP Simulator, dual lumen hose w/ NIBP test fixture cuff (9100-3072-TF) or equivalent

Note:

If ECG simulator is connected to the device, either disconnect ECG simulator or
disable the Smart Cuff feature in the Supervisor Menu.

Setup

Attached the hose connected to the T-fitting of the NIBP test fixture cuff to the pressure port of
the NIBP simulator. Set simulator to Static Pressure.

Do This...

Observe This...

Pass/Fail

20.1 Turn on the X Series unit.

Note: If the unit is in AED mode, press
the Manual quick access key to go into
manual mode.

20.2 Connect a dual lumen hose from the NIBP

simulator to the NIBP port of the X
Series.

20.3 On the X Series unit, do the following to

enter the NIBP Test control panel:

Press the More (

) then the Setup

(

) quick access keys.

Use the navigation keys to select
Supervisor>Service>Device
Tests
>NIBP Test.

Note: You need a supervisor passcode to
enter the Supervisor menu.

20.4 From the X Series NIBP Test control

panel, use the navigation keys to select:

Close Valve

Turn Pump On

20.5 Monitor the pressure reported by the

external pressure standard.

20.6 Once the pressure reaches the target

pressure of 50 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.

20.7 Allow the system to stabilize for at least

10 seconds.

20.8 Verify that the primary transducer agrees

with the reading on the NIBP simulator.

Reading should be within
3mmHg.

o

o

20.9 Verify that the safety transducer agrees

with the reading on the NIBP simulator.

Reading should be within
4mmHg.

o

o