Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

9

2.0 Operational Tests

The X Series unit comes equipped with routine operational tests that can be performed
periodically. With the exception of the Physical Inspection of the Unit test, these tests contain
instructions within them that provide guidance during the tests.

The operational tests on the unit can be accessed by doing the following:

1. Turn on the X Series unit.

Note:

If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to select the four digits in the Manual Mode pass code.
Press SAVE when you are finished. If the unit has not been configured to enter a pass
code, the message Exit to Manual Mode is displayed. Use the navigation keys to select
Yes.

2.

Press the More (

) then the Setup (

) quick access keys.

3.

Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests
are displayed. Use the navigation keys to select one of the tests.

Note:

You need a supervisor passcode to enter the Supervisor menu.