Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

33

16.0 EtCO

2

Calibration (required for annual calibration)

Equipment

5% CO

2

calibration gas cylinder with CO

2

sample line (kit available from Scott Medical

Products, PN 4653ORF-BD)

Do This...

Observe This...

Pass/Fail

16.1

Turn on the X Series unit.

16.2

On the X Series unit, select
Supervisor>Service>Diagnostic>CO

2

>

Calibration to enter the CO

2

Calibration

control panel.

16.3

Connect the X Series unit to the 5% CO

2

calibration gas cylinder using the sample
line included with the calibration kit.

16.4

From the CO

2

Calibration Menu, select

Calibration.

16.5

Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.

16.6 Continue to apply the calibration gas until the

status in the CO

2

Calibration control panel

displays: CALIBRATION GAS CAN BE
REMOVED
.

When the calibration is complete, the
status in the CO

2

Calibration control panel

displays: CALIBRATION DONE-OK

The calibration date will be updated upon
exiting the CO

2

Test control panel.

o

o

o

o