32

9650-001356-01 Rev. D

15.5 Set the simulator to 95%

SpO

2

and 230

BPM.

Verify that after a brief period of
initialization,
The

SpO

2

numeric displays an

SpO

2

value between 91.3% and

98.8%.
The

SpCO

numeric displays an

SpCO

value between -.5 and 7.5.

The

SpMet

numeric displays an

SpMet

value between -.5 and 3.5.

Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.

o

o

o

o

o

o

o

o

o

15.6 Set the simulator to 95%

SpO

2

and 50

BPM.

Verify that after a brief period of
initialization,
The

SpO

2

numeric displays an

SpO

2

value between 91.3% and

98.8%.
The

SpCO

numeric displays an

SpCO

value between -.5 and 7.5.

The

SpMet

numeric displays an

SpMet

value between -.5 and 3.5.

Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.

o

o

o

o

o

o

o

o

o

15.7 Disconnect the

SpO

2

simulator.

Do This...

Observe This...

Pass/Fail/N/A