beautypg.com

0 leads test, Leads test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 21

background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

15

3.0 Leads Test

Equipment

3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer

Note:

Test each cable separately.

Test Setup

Connect the lead wires appropriate for each test to the

Fluke Impulse 7000DP Analyzer

or

equivalent, then connect the ECG cable to the X Series unit.

Do this...

Observe this...

Pass/Fail

3.1 Turn on the X Series unit.

Note: Make sure the lead is not
displayed.

No ECG LEAD FAULT message
displayed.

o

o

3.2 Disconnect one lead from the

simulator.

The ECG LEAD FAULT message displays
within 3 seconds (if configured).

o

o

3.3 Reconnect the lead. Repeat step 2

with the remaining leads.

Wait for ECG LEAD FAULT message to
clear from the display (if configured).

o

o

3.4 Repeat steps 2 and 3 for 5 lead and 12

lead cables.

Note: If heart rate alarm sounds, on the
X Series unit:

Press the More quick access key (

) key

to access the second set of quick access
keys.

Press the Alarm quick access key (

).

Press the Alarm Cancel quick access key

(

).

o

o