0 pacer test, Pacer test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 33
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
27
11.0 Pacer Test
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent)
Do This...
Observe This...
Pass/Fail
11.1 On the Analyzer, select the PACER
function per test equipment instructions.
11.2 On the Analyzer, press F2 to select
PACER JACKS, then press F5 (Done) to
measure pulse.
11.3 Connect the MFC connector to the
Analyzer.
11.4 On the X Series unit, apply power to the
unit using a battery or an AC adapter.
11.5 On the X Series unit, press the PACER
button.
11.6 On the X Series unit, use the navigation
keys to do the following:
•Set Rate to 30.
•Set Mode to Fixed.
•Set Output to 10.
•Select Start Pacer.
o
o
11.7 On the Impulse 7000 DP Analyzer,
record the following:
•pulse rate
•pacer amplitude output
•pacer pulse width
•Pacer Rate PPM
(28.5ppm-31.5ppm)
•Pacer amplitude output
(5mA-15mA)
•Pacer pulse width (38ms to 42ms)
Note: If an ECG FAULT message
displays when leads are not
attached, it is normal.
11.8 On the X Series unit, use the navigation
keys to do the following:
•Set Rate to 180.
•Set Output to 140.
o
o
11.9 On the Impulse 7000 DP Analyzer,
record the following:
•pulse rate
•pacer amplitude output
•pacer pulse width
•Pacer Rate PPM
(162ppm-198ppm)
•Pacer amplitude output
(133mA-147mA)
•Pacer pulse width (38ms to 42ms)
11.10 Select Turn Pacer Off.