beautypg.com

0 pacer test, Pacer test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 33

background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

27

11.0 Pacer Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent)

Do This...

Observe This...

Pass/Fail

11.1 On the Analyzer, select the PACER

function per test equipment instructions.

11.2 On the Analyzer, press F2 to select

PACER JACKS, then press F5 (Done) to
measure pulse.

11.3 Connect the MFC connector to the

Analyzer.

11.4 On the X Series unit, apply power to the

unit using a battery or an AC adapter.

11.5 On the X Series unit, press the PACER

button.

11.6 On the X Series unit, use the navigation

keys to do the following:

•Set Rate to 30.

•Set Mode to Fixed.

•Set Output to 10.

•Select Start Pacer.

o

o

11.7 On the Impulse 7000 DP Analyzer,

record the following:

•pulse rate

•pacer amplitude output

•pacer pulse width

•Pacer Rate PPM
(28.5ppm-31.5ppm)

•Pacer amplitude output
(5mA-15mA)

•Pacer pulse width (38ms to 42ms)

Note: If an ECG FAULT message
displays when leads are not
attached, it is normal.

11.8 On the X Series unit, use the navigation

keys to do the following:

•Set Rate to 180.

•Set Output to 140.

o

o

11.9 On the Impulse 7000 DP Analyzer,

record the following:

•pulse rate

•pacer amplitude output

•pacer pulse width

•Pacer Rate PPM
(162ppm-198ppm)

•Pacer amplitude output
(133mA-147mA)

•Pacer pulse width (38ms to 42ms)

11.10 Select Turn Pacer Off.