Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

25

9.0 Advisory Test

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent), ECG simulator

Do This...

Observe This...

Pass/Fail

9.1

Connect the multifunction cable (MFC) to
the Analyzer.

9.2

Turn on the X Series unit and connect the
pads.

The unit displays Pads on the top
left of the screen.

o

o

9.3

On the Analyzer, select VF (ventricular
fibrillation), then press the Analyze button.

The X Series unit displays:

ANALYZING

SHOCK ADVISED

o

o

9.4

On the X Series unit, press the CHARGE
button to activate, then press the CHARGE
button again.

The unit charges to the selected
energy level.

o

o

9.5

On the X Series unit, press the Shock button. Verify that the shock count is

displayed at the top of the screen.

o

o

9.6

On the Analyzer, select NSR (normal sinus
rhythm), then press the Analyze button.

The X Series unit displays:

ANALYZING

NO SHOCK ADVISED

o

o