Maintenance tests, 0 physical inspection of the unit, Physical inspection of the unit – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 14
C
HAPTER
1
M
AINTENANCE
T
ESTS
8
9650-001356-01 Rev. D
Maintenance Tests
1.0 Physical Inspection of the Unit
Observe this...
Pass/Fail
1.1
Housing
Is the unit clean and undamaged?
o
o
1.2 Does the unit show signs of excessive wear?
o
o
1.3 Does the handle work properly?
o
o
1.4 Does the recorder drawer open and close properly?
o
o
1.5 Are input connectors clean and undamaged?
o
o
1.6 Are there any cracks in the housing?
o
o
1.7 Does the front panel have any damage or cracks?
o
o
1.8 Are there any loose housing parts?
o
o
1.9 Paddles (if applicable)
Do the adult and pedi plates have major scratches or show signs of damage?
o
o
1.10 Do the adult shoes slide on and off easily to expose the covered pedi plates?
o
o
1.11 Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
o
o
1.12 Cables
Are all cables free of cracks, cuts, exposed or broken wires?
o
o
1.13 Are all bend/strain reliefs undamaged and free of excessive cable wear?
o
o
1.14 Battery
Is the X Series battery fully charged?
o
o
1.15 Is the battery seated in the battery well correctly?
o
o
1.16 Is battery housing free of cracks and damage?
o
o
1.17 Are connectors free of damage and excessive wear?
o
o