8.0 Shock Test

WARNING!

Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.

Keep hands and all other objects clear of the multi-function cable connections and
defibrillator analyzer when discharging the defibrillator.

Before you discharge the defibrillator, warn everyone near the equipment to STAND
CLEAR.

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent), stop watch

Do This...

Observe This...

Pass/Fail

8.1

Turn on the Analyzer.

8.2

On the Analyzer, select the DEFIB function.

8.3

On the Analyzer, select ENERGY.

8.4

Connect the X Series MFC connector to the
pads connection.

8.5

On the X Series unit, apply power to the unit
using a battery or an AC adapter.

8.6

Connect the MFC cable to the X Series unit.

8.7

Turn on the X Series unit, and press the

Leads quick access key

(

)

to select

PADS

.

8.8

On the X Series unit, set energy level to 5J
by pressing the Energy Select DOWN arrow.

8.9

On the X Series unit, press the CHARGE
button.

The unit charges to 5J, the red
Shock button illuminates and a
ready tone sounds.

o

o

8.10 On the X Series unit, press the Shock button.

8.11 On the Analyzer, observe and record the

energy measurement.

Energy is delivered at 5J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24
.

o

o

8.12 On the X Series unit, set the energy level to

50J by pressing the Energy Select UP arrow.

8.13 On the X Series unit, press the CHARGE

button.

The unit charges to 50J, the red
Shock button illuminates and a
ready tone sounds.

o

o

I, II,