0 synchronized cardioversion test, Synchronized cardioversion test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 32

background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

26

9650-001356-01 Rev. D

10.0 Synchronized Cardioversion Test

Equipment

Fluke Impulse 7000DP Analyzer (or equivalent)

Test Setup

Connect the cable to the defibrillator analyzer.

Select Sync mode on analyzer.

Input 1mV ECG signal at 60 BPM.

Do this...

Observe this...

Pass/Fail

10.1 Verify that the X Series unit is turned

on and the

PADS

lead is selected.

10.2 Press the SYNC softkey on the

defibrillator. Enter the synchronized
cardioversion timing test mode on the
defibrillator analyzer.

Sync appears on display.

Sync markers display on the monitor.
The sync marker appears as an S over
the ECG R-wave peaks on strip chart
and display.

o

o

10.3 Select 200J.

10.4 Press the CHARGE button. When the

Shock button illuminates, press and
hold it.

Observe that the R-wave to shock delay
(sync delay) is less than 60 milliseconds
on the analyzer display.

Defibrillator discharges.

o

o