Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
31
15.0 SpO
2
, SpCO, and SpMet Test
(for SpO
2
feature with SpCO, or SpCO and SpMet)
Note:
Use this test if your unit has SpO
2
and SpCO, or SpO
2
, SpCO and SpMet. If you do not
have the SpMet feature on the unit, disregard the SpMet values in the test below.
Equipment
Fluke Index 2 SpO
2
Simulator (or equivalent), Masimo Rainbow 3 ft BioTek Test Sensor
(Masimo P/N 3379) or Masimo Rainbow 3 ft Direct Connect Optical Test Sensor
Test Setup
Disconnect any ECG cable or IBP probes from the X Series unit.
Do This...
Observe This...
Pass/Fail/N/A
15.1 Power on the X Series unit without the
SpO
2
sensor attached.
Verify that the
SpO
2
numeric
displays INITIALIZING, then
goes blank.
15.2 Connect the
SpO
2
sensor to the X
Series unit.
Verify that
SpO
2
numeric
displays SEARCHING.
15.3 Set the simulator to 75%
SpO
2
and 50
BPM. Attach the
SpO
2
sensor to the
simulator.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 72.1% and
78%.
The
SpCO
numeric displays an
SpCO
value between 13.5 and
21.5.
The
SpMet
numeric displays an
SpMet
value between 5.5 and
9.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
o
o
o
o
o
o
o
o
o
15.4 Set the simulator to 75%
SpO
2
and 230
BPM.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 72.1% and
78%.
The
SpCO
numeric displays an
SpCO
value between 12.8 and
20.8.
The
SpMet
numeric displays an
SpMet
value between 5.2 and
9.2.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
o
o
o
o
o
o
o
o
o