Before you begin the maintenance tests – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 12

background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

6

9650-001356-01 Rev. D

This chapter describes the following maintenance tests:

1.0 Physical Inspection of the Unit

2.0 Operational TestsOperational Tests

3.0 Leads Test

4.0 Heart Rate Display Test

5.0 Heart Rate Alarm Test

6.0 Power Supply Test

7.0 Leakage Current Test

8.0 Shock Test

9.0 Advisory Test

10.0 Synchronized Cardioversion Test

11.0 Pacer Test

12.0 Printer Test

13.0 Treatment Report Test

14.0 SpO2 Test (for SpO

2

feature without SpCO and SpMet)

15.0 SpO2, SpCO, and SpMet Test (for SpO

2

feature with SpCO, or SpCO and SpMet)

16.0 EtCO2 Calibration (required for annual calibration)

17.0 EtCO2 Calibration Check

18.0 EtCO2 Test

19.0 IBP Test

20.0 NIBP Calibration Check

21.0 NIBP Leak Test

22.0 Deflation Test

23.0 Inflation Test

24.0 NIBP Test

25.0 Temperature Test

26.0 Wi-Fi Test (Optional)

27.0 Audio Recording Test (Optional)

Before You Begin the Maintenance Tests

Assemble the tools listed in following section, “Equipment You Need to Perform the

Maintenance Tests.”

Keep an extra fully charged X Series battery available.

Schedule an hour to conduct the entire maintenance test.

Perform the tests in the order presented.

Perform all the steps of each test procedure.

Complete all the steps of the procedure before evaluating the test results.