beautypg.com

0 deflation test, Deflation test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 47

background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

41

22.0 Deflation Test

Equipment

Dual lumen NIBP hose, stop watch, NIBP simulator

Do This...

Observe This...

Pass/Fail

22.1 From the X Series NIBP Test control

panel, use the navigation keys to select:

Close Valve

Turn Pump On

22.2 Monitor the pressure reported by the

NIBP Test control panel or the external
pressure standard, and select Turn Pump
Off
as soon as the pressure reaches the
target pressure of 265 +/- 5 mmHg.

22.3 Allow the system to stabilize for 5

seconds.

22.4 Use the navigation keys to select Open

Valve and immediately start the timer.

22.5 Stop the timer as soon as pressure drops

below 15 mmHg.

The pressure should drop within 10
seconds.

o

o