0 heart rate alarm test, Heart rate alarm test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 23
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
17
5.0 Heart Rate Alarm Test
Equipment
Fluke Impulse 7000DP Analyzer
Do this...
Observe this...
Pass/Fail
5.1
Press the Leads quick access key
(
)
on the X Series unit and
select lead II.
II displays next to the Leads quick
access key.
o
o
5.2
On the X Series unit:
•
Press the More quick
access key (
) key to
access the second set of
quick access keys.
•
Press the Alarm quick
access key (
).
•
Press the Limits softkey.
The Alarm Settings menu displays.
o
o
5.3
On the X Series unit, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:
•
Set the Upper Limit Enable
to ON.
•
Set the Lower Limit
Enable to ON.
•
Set the Lower limit to 35.
•
Set the Upper limit to 150.
Use the navigation keys to exit the
Alarm Settings menu.
5.4
On the X Series unit, press the
Alarm Cancel quick access key
(
).
The red alarm suspension timer
displays at the top of the screen next
to the message area.
5.5
Remove a lead wire from the
ECG
simulator
.
T
he ECG LEAD FAULT message
displays.
o
o
5.6
On the X Series unit, press
.
The Alarm Suspend key flashes
.
5.7
Reattach ECG Lead wire to
ECG
simulator
.
The ECG LEAD FAULT message no
longer displays.
o
o
5.8
Set simulator to 30 BPM or lower.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM message displays.
5.9
On the X Series unit, press
.
The Alarm Suspend key flashes.
The
red alarm suspension timer displays
at the top of the screen next to the
message area.
o
o
5.10 Set the simulator to 120 BPM.
5.11 On the X Series unit, press
.
Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.
o
o
I, II,
III...