background image

Disassembling the Main Unit

9650-001356-01 Rev. D

X Series Service Manual

67

3

Remove ten (10) screws: four
(4) on the top, four (4) on the
bottom, and two (2) in the
middle in the illustrated
areas. Take care to keep the
two parts of the unit together.

Note: During reassembly,
torque screws to 6 in-lbs.

4

While holding the two parts of
the unit together, place the
unit vertically on the mat (feet
down).

5

Separate the top part of the
unit by creating a small gap,
and remove any screws that
are in the way.