Spo2 test (for spo, 0 spo2 test (for spo, Feature without spco and spmet) – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 36: 0 spo, Test (for spo
C
HAPTER
1
M
AINTENANCE
T
ESTS
30
9650-001356-01 Rev. D
14.0 SpO
2
Test (for SpO
2
feature without SpCO and SpMet)
Note:
Perform this test if your unit has SpO
2
only. See the following procedure to test units
with SpO
2
, SpCO, and SpMet.
Equipment
Fluke Index 2 SpO
2
Simulator (or equivalent), Masimo
®
Rainbow
®
SpO
2
finger probe
Test Setup
Disconnect any ECG cable or IBP probes from the X Series unit.
Do This...
Observe This...
Pass/Fail
14.1 Power on the X Series unit without the
SpO
2
sensor attached.
Verify that the
SpO
2
numeric
displays INITIALIZING, then
goes blank.
14.2 Connect the
SpO
2
sensor to the X Series unit. Verify that
SpO
2
numeric
displays SEARCHING.
14.3 Set the simulator to 70%
SpO
2
and 60 PPM.
Attach the
SpO
2
sensor to the simulator.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
14.4 Set the simulator to 80%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
14.5 Set the simulator to 90%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
o
o
o
o
14.6 Disconnect the
SpO