ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 61

X Series Error Messages
9650-001356-01 Rev. D
X Series Service Manual
55
NIBP AIR LEAK
Major air leak or could not 
reach target pressure.
Attempt to identify 
problems with the hose 
and/or cuff by doing the 
following:
•
Check the hose 
connection to the 
device. Make sure that 
the hose is tightly 
screwed onto the NIBP 
connector.
•
Ensure that the cuff is 
appropriately wrapped 
and is not too loose.
•
Inspect the cuff and 
hose for leaks. Inspect 
the O-ring on the NIBP 
hose connector. 
Consider replacing the 
cuff and/or hose.
If the problem persists, 
perform the following 
procedure: “NIBP Leak 
Test” on page 40:
Inspect internal NIBP 
tubing for leaks.
NIBP ARTIFACT
Unable to determine blood 
pressure. Check 
connections; restrict patient 
movement.
Attempt to identify and 
eliminate the source of 
artifact. If problem persists, 
replace the NIBP module.
NIBP COMMUNICATION 
FAILURE
There is a communication 
problem with the NIBP 
module.
Ensure that the J32 
connection is secure. If 
error persists, replace the 
NIBP module.
NIBP EQUIPMENT CHECK
Device is unable to obtain a 
reading.
Check the hose and cuff 
connections. If problem 
persists, perform one or 
more of the following 
procedures in Chapter 1:
•
NIBP Leak Test
•
NIBP Deflation Test
•
NIBP Inflation Test
NIBP KINKED HOSE
The cuff inflation was too 
rapid. There may be kinked 
tubing or another air 
obstruction.
Check the external NIBP 
hose for kinks. If problem 
persists, check the routing 
of the NIBP tubing inside 
the device.
Error Message
Explanation
Technical Action