beautypg.com

ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 29

background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

23

8.14 On the X Series unit, press the Shock button.

8.15 On the Analyzer, observe and record the

energy measurement.

Energy is delivered at 50J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24
.

o

o

8.16 On the X Series unit, set energy level to 100J

by pressing the Energy Select UP arrow.

8.17 On the X Series unit, press the CHARGE

button.

The unit charges to 100J, the
Shock button illuminates and a
ready tone sounds.

o

o

8.18 On the X Series unit, press the Shock button.

8.19 On the Analyzer, observe and record the

energy measurement.

Energy is delivered at 100J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.

o

o

8.20 On the X Series unit, set energy level to 200J

by pressing the Energy Select UP arrow.

8.21 On the X Series unit, press the CHARGE

button.

The unit charges to 200J, the red
Shock button illuminates and a
ready tone sounds.

o

o

8.22 On the X Series unit, press the Shock button.

8.23 On the Analyzer, observe and record the

energy measurement.

Energy is delivered at 200J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.

o

o

8.24 On the X Series unit, press the CHARGE

button.

The unit charges to 200J, the red
Shock button illuminates, and a
ready tone sounds.

o

o

8.25 When the X Series unit is charged, start the

stop watch.

Verify that after 50 seconds the
tone changes and within 10
seconds the unit does an internal
discharge and displays
DISARMED TIMEOUT.

o

o

8.26 Turn off the X Series unit.

Do This...

Observe This...

Pass/Fail