ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 29
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
23
8.14 On the X Series unit, press the Shock button.
8.15 On the Analyzer, observe and record the
energy measurement.
Energy is delivered at 50J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24.
o
o
8.16 On the X Series unit, set energy level to 100J
by pressing the Energy Select UP arrow.
8.17 On the X Series unit, press the CHARGE
button.
The unit charges to 100J, the
Shock button illuminates and a
ready tone sounds.
o
o
8.18 On the X Series unit, press the Shock button.
8.19 On the Analyzer, observe and record the
energy measurement.
Energy is delivered at 100J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.
o
o
8.20 On the X Series unit, set energy level to 200J
by pressing the Energy Select UP arrow.
8.21 On the X Series unit, press the CHARGE
button.
The unit charges to 200J, the red
Shock button illuminates and a
ready tone sounds.
o
o
8.22 On the X Series unit, press the Shock button.
8.23 On the Analyzer, observe and record the
energy measurement.
Energy is delivered at 200J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.
o
o
8.24 On the X Series unit, press the CHARGE
button.
The unit charges to 200J, the red
Shock button illuminates, and a
ready tone sounds.
o
o
8.25 When the X Series unit is charged, start the
stop watch.
Verify that after 50 seconds the
tone changes and within 10
seconds the unit does an internal
discharge and displays
DISARMED TIMEOUT.
o
o
8.26 Turn off the X Series unit.
Do This...
Observe This...
Pass/Fail