ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 13
X Series Overview
9650-001356-01 Rev. D
ZOLL X Series Service Manual
7
Equipment You Need to Perform the Maintenance Tests
The equipment listed below is used in the maintenance procedures in this chapter. You can
substitute an equivalent device; however, not all simulators and analyzers will produce the
same results. Be sure to follow the manufacturer’s recommendations for conducting the
maintenance tests.
•
Fluke Impulse 7000DP Analyzer, with appropriate interface adapter (Fluke Biomedical
BP Pump/Dynatech Nevada Inc. part number 3010-0378)
•
ECG simulator
•
Fluke Index 2 SpO
2
simulator
•
IBP/Temp simulator
•
NIBP simulator
•
5% CO
2
calibration gas cylinder with CO
2
sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
•
CO
2
sampling line airway adapter
•
Stop watch
•
Battery Current Test Fixture 9100-3055-TF
•
Agilent E3633A Power Supply
•
Agilent 34401 Digital Multi-meter
Accessories You Need for the X Series Options Maintenance Tests
•
3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
•
Dual lumen NIBP hose
•
IBP cable
•
SpO
2
cable
and sensor
•
EtCO
2
cable
and sensor
•
Paddles
•
Printer paper
•
Battery
•
Auxiliary power source
•
USB thumb drive
•
PC w/ Windows media player and speakers