Quickstart, English – Burkert Type 8175 User Manual
Page 76

8
8175
ENGLISH
QUICKSTART
To access the CALIBRATION MENU simutaneously press
ENTER
for 5 seconds.
-
The ‘Enter‘ key can be locked to avoid accidental or unauthorised access - section 3.3.
-
Reference can additionally be made to the menu guide - section 4.2.
-
It is important to exit the menu calibration after entering the desired distance units and
re-enter the menu to configure the volume units.
2.2.3
Programming - Volume
The actions which are highlighted in grey
must be fully completed for accurate
measurement.
Choose the
desired language
Select the desired
distance units
Quit this menu and
re-enter
Select the desired
volume units
Select the desired
filter
Enter the
Sound velocity
Perform a Teach-In
procedure
Enter the current
output range
Configure relays
1 & 2 ( if available )
See section 4.4.1
See section 4.4.2
See section 4.4.2
See section 4.4.3
See section 4.4.4
See section 4.4.6
See section 4.4.7
See section 4.4.8
Measurement or Simulation (optional)
See the next page.
It is important to quit this
calibration menu and re-
enter after entering the
distance units to ensure
correct measurement.
Choose filters 0-2 if no echo filtering
of obstructing obstacles is required
and filters 3-9 if echo filtering is
required.
331.4 m/s within the air with a devia-
tion of 0.59m/s/°C.
For ex. 4mA -> 1.2m
3
and 20mA -> 10.6m
3