Calibration menu, English – Burkert Type 8175 User Manual
Page 114

46
8175
ENGLISH
CALIBRATION MENU
4.4.7 Output
Current
Within this option the measuring range can be defined corresponding to the output current
of 4-20 mA.
-
The beginning of the
measuring range might be larger than the end creating an
inverted signal, e.g. 0.5 to 6m corresponds to 20-4 mA.
-
The engineering units and decimal point selected will be valid within this option.
-
The basic setting of the span and offset can be checked within the test menu.
0......9
ENTER
0......9
ENTER
ENTER
Enter the beginning of the
measuring range
Enter the end of the measuring range
-
In case of signal loss the device will emit an error signal of 22mA.
-
The diagram below shows the type of relationship between the 4-20mA ouptut and the
associated measuring range.
CURRENT
RELAY
4=00.00
4=02.00
20=00.00
20=12.00
2
20
mA
4
12
m