4.5.3 Temperature
adjustment
The 8175 level transmitter posseses a temperature probe within the sensor. The value from
this sensor can be influenced by an offset entered within this option. To enter a desired
offset press
ENTER
when «T ° ADJUST» is displayed within the test menu and enter the value.
-
The offset range is + or - 15°C.
-
The temperature adjustment remains active until another temperature value is entered.
4.5.4
STRENGTH OF RETURN SIGNAL
This function allows the power of the ultrasonic echo received by the sensor to be viewed.
When the optimum signal strength is 90 dB the transmitter has a measuring range of 7m for
the remote version and 10m fo compact version. The strength of this signal is maximum if
the temperature of the gas and liquid are equal or the surface of the fluid is calm.
This measure is useful to determine the best orientation of the transmitter during installation.
The best orientation is reached when the signal strength is at the highest value possible.
0......9
ENTER
ENTER
Enter the temperature offset in
°C if chosen unit=cm or m and
°F if chosen unit=foot or inch
+
0.0 ° C
+
1.2 ° C
T ° ADJUST
SIGNAL
SIMUL
90.00 dB
SIGNAL