Specifications, English, 1 specifications – Burkert Type 8175 User Manual
Page 126
58
8175
ENGLISH
SPECIFICATIONS
6.1 SPECIFICATIONS
Specifications in relation to the process
Measurement
Measurement type
Level, distance or volume measurement
Measurement range
0.30 to 10 meters compact version and
0.30 to 7 meters wall and panel versions
Measurement range in the
following conditions:
Measure of the fluid without foam,
Tambient = 20°C ; Tfluid = 20°C
Ambient pressure value : atmopheric pressure
Accuracy
+/- 0.25% of the full scale
+/- 0.15% after calibration with Teach-In
Resolution
+/-
3mm
Full beam width
8 degrees conical
Pulse rate
8 pulses per second
Blocking distance
30 cm from the base of the sensor
Installation
Pressure rating
2 bar at 25°C maximum
Mounting threads
G 2‘‘ or NPT 2‘‘
Fluid temperature
-40 to +80°C / -40 to +176°F
Gasket material
FPM or EPDM (option)
Specifications in relation to the outputs
Electrical connection
Power
supply
18-32VDC
115/230 VAC 50-60Hz (+/- 10% VAC only)
Consumption
200mA
maximum
Proportional ouput
Output type
Current output of 4-20mA (22mA error signal)
Accuracy
+ / - 2%
Wiring
Sinking or sourcing
Maximum load
1300 Ohms at 32V
1000 Ohms at 24V
550 Ohms at 18V
Output calibration
4-20mA or inverted to 20-4mA adjustable
Relay ouput
Output type
2 relays (3A), normally open
Load
DC : 250V, 3A
AC : 250V, 3A
Life expectancy
100 000 cycles (minimum)
Switching
Thresholds, hysteresis and delay programmable