beautypg.com

Calibrating a ph or redox sensor – Burkert Type 8619 User Manual

Page 93

background image

91

Adjustmentandcommissioning

8.11.8. calibrating a ph or redox sensor

Refer to chap. 8.9 to access Calibration menu.

Calibration

pH auto calib.

Calib. log

Mx:pH/ORP

pH manual calib.

1st point

2nd point?

Rinse

2nd point

pH Cal.Result

YES

NO

pH calibration data

Last

Iso potential

VALUE

Iso potential

Zero

ENTERING

Slope

ENTERING

Iso pH

ENTERING
ENTERING

ORP calibration

1st point

ORP Cal.Result

ORP calib. data

Offset

ENTERING

ENTERING

Calib. interval

READING

Interval

Use one of the following 3 methods to calibrate a pH sensor:

PH AUTO CALIBRATION: Automatically calibrate the pH sensor; first choose the buffer solution used in the
menu "Parameters -> Mx:pH/ORP -> Buffer". The multiCELL automatically recognizes the pH of the buffer
used. When a sensor is calibrated automatically, the most recent calibration date is updated (function "LAST" in
the sub-menu "CALIBRATION INTERVAL" below).

PH MANUAL CALIBRATION: Calibrate the pH sensor using a 1- or 2-point procedure. See details on p. 84.
When a sensor is calibrated manually, the most recent calibration date is updated (function "LAST" in the sub-
menu "CALIBRATION INTERVAL" below).

PH CALIBRATION DATA: Enter the zero values and the slope indicated on the certificate for the pH sensor, if
provided. This entry does not update the most recent calibration date (function "LAST" in the sub-menu "CALI-
BRATION INTERVAL" below).

Use one of the following 2 methods to calibrate a redox sensor:

ORP CALIBRATION: Calibrate the redox sensor using a 1-point procedure. See details on page 94.

ORP CALIBRATION DATA: Enter the offset value indicated on the certificate for the redox sensor, if provided.

CALIBRATION INTERVAL: Read the date of the most recent manual or automatic calibration and enter the
periodicity of calibration, in days: each time a calibration is due, the multiCELL generates a "maintenance" event,
signalled on the display by the icon , and a "warning" event, signalled on the display by the icon

. Configure

"0000 days" in order not to use the function.

• The "warning" event may be associated with one and/or other digital outputs (see chap. 8.10.22).

• See also "If you encounter problems", chap. 9.3

Type 8619

english