Adjustmentandcommissioning

CALIBRATION LIMITS: Enter the ranges outside of which a warning or error message is generated during
calibration:

• PH ZERO:

- WARNING HIGH: Enter the pH0 value above which a warning message is displayed during calibration of the pH sensor.

- WARNING LOW: Enter the pH0 value below which a warning message is displayed during calibration of the pH sensor.

- ERROR HIGH: Enter the pH0 value above which an error message is displayed during calibration of the pH sensor.

- ERROR LOW: Enter the pH0 value below which an error message is displayed during calibration of the pH sensor.

• PH SLOPE:

- WARNING HIGH: Enter the slope value above which a warning message is displayed during calibration of the pH sen-

sor.

- WARNING LOW: Enter the slope value below which a warning message is displayed during calibration of the pH sensor.

- ERROR HIGH: Enter the slope value above which an error message is displayed during calibration of the pH sensor.

- ERROR LOW: Enter the slope value below which an error message is displayed during calibration of the pH sensor.

• ORP OFFSET:

- WARNING HIGH: Enter the oxidation reduction potential value above which a warning message is displayed during

calibration of the redox sensor.

- WARNING LOW: Enter the oxidation reduction potential value below which a warning message is displayed during cali-

bration of the redox sensor.

- ERROR HIGH: Enter the oxidation reduction potential value above which an error message is displayed during calibra-

tion of the redox sensor.

- ERROR LOW: Enter the oxidation reduction potential value below which an error message is displayed during calibration

of the redox sensor.

Type 8619

english