Adjustmentandcommissioning
The multiCELL associates the two calibrated values to the range of the selected
analogue input signal (slope and offset).
→
Immerse the probe of the sensor (for insatne a pH sensor) in the first buffer
solution: the multiCELL displays the solution measured value.
→
Enter the value of the buffer solution (indicated on the bottle).
→
Wait until the measured value stabilises.
Calibration
Mx:Inputs
PV calibration
OK
F4
1st point
7.001 pH
7.000 pH
Calibrate using a
2nd point?
NO
2nd point
10.001 pH
10.000 pH
→
Rinse the probe.
→
Immerse the clean probe in the second buffer solution: the multiCELL dis-
plays the measured value of the solution.
→
Enter the value of the second buffer solution (indicated on the bottle).
→
Wait until the measured value stabilises.
OK
F4
YES
The multiCELL adjusts the
offset.
AI1/AI2
Solution measured value.
Calibration result
Offset: -3,498 pH
Slope:0,876 pH/mA
SAVE
F4
Figure 68 :
Example for the calibration of an analogue input with respect to the pH measured by an instrument having a
4-20 mA output
Type 8619
english