beautypg.com

Wiring the "cond" conductivity module – Burkert Type 8619 User Manual

Page 31

background image

29

Installationandwiring

7.3.6.

Wiring the "conD" conductivity module

Conductivity

sensor

C+

P+

P-

C-

GD

FE

SE

TS

TS

1 2 3 4 5 6 7 8 9

Temperature

sensor

FE = functional earth

Figure 18 :

Wiring a resistive conductivity cell with 2 electrodes and a Pt100 or Pt1000 temperature sensor in a conduc-
tivity module

C+

P+

P-

C-

GD

FE

SE

TS

TS

1

3

2

8220

1 2 3 4 5 6 7 8 9

FE = functional earth

Figure 19 :

Connection example for a 8220 conductivity sensor

C+

P+

P-

C-

GD

FE

SE

TS

TS

1 2 3 4 5 6 7 8 9

Temperature

sensor

FE = functional earth

Figure 20 :

Wiring a resistive conductivity cell with 4 electrodes and a Pt100 or Pt1000 temperature sensor in a conduc-
tivity module

Type 8619

english