Setting the parameters of a conductivity module, Refer to chap. 8.9 to access parameters menu, If "temperature" = "manual – Burkert Type 8619 User Manual
Page 83: If "temp. calib" = "manual, English
81
Adjustmentandcommissioning
8.10.24. setting the parameters of a conductivity module
Refer to chap. 8.9 to access Parameters menu.
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
Mx:Conductivity
Temp. adjust
RTD
Temp. calib
Parameters
PT100
None
PT1000
ENTERING
Concentration
H2SO4(32-84%)
H2SO4(0-30%)
HNO3(0-30%)
None
Temperature
Auto
Manual
Auto
Manual
Calibration Buffer
5µS
100mS
H2SO4(92-99%)
HCl(0-18%)
HNO3(35-96%)
NaOH(0-14%)
HCl(22-39%)
NaCl(0-26%)
NaOH(18-50%)
Cell
2 electrodes
4 electrodes
Temp.Comp.
None
UPW-NaCl
Linear
EN 27888
UPW
Coef.
ENTERING
if "Temp.Comp." = "Linear"
if "Temperature" = "Auto"
°C
ENTERING
if "Temperature" = "Manual"
°C
ENTERING
if "Temp. calib" = "Manual"
15µS
100µS
1413µS
1)
USP alarm
ENTERING
Concentr.table
1)
This menu is available as an option: see chap. 8.10.4.
CELL: Choose the type of conductivity cell, either with 2 or 4 electrodes, connected to the module.
RTD: Choose the type of temperature sensor connected to the module.
TEMPERATURE: Choose the value of the temperature used in the process:
• choose "Auto": the temperature of the fluid is measured by the sensor
• choose "Manual": enter the value of the process temperature (in °C) in the next field, e.g. when no temperature
sensor is connected to the module.
Type 8619