Calibrating the analogue inputs – Burkert Type 8619 User Manual
Page 87

85
Adjustmentandcommissioning
8.11.4. calibrating the analogue inputs
This calibration does not replace the calibration of the measuring sensor.
Refer to chap. 8.9 to access Calibration menu.
Mx:Inputs
Voltage calibr.
AI1/AI2
PV calibration
1st point
2nd point?
2nd point
Calibration result
YES
NO
Calibration
Current calibr.
An analogue input, AI1 or AI2, can be calibrated:
• either with respect to the measured process variable (function "PV calibration"). See Figure 68.
• or with respect to the current or voltage source connected to the input. This calibration is used to precisely
adjust the bounds of the analogue input to the bounds of the connected instrument.
Type 8619
english
See also other documents in the category Burkert Sensors:
- Type 1062 (112 pages)
- Type 1050 (4 pages)
- Type 8750 (64 pages)
- Type 8750 (82 pages)
- Type 8681 (40 pages)
- Type 8681 (90 pages)
- Type 8798 (2 pages)
- Type 8791 (4 pages)
- Type 8798 (106 pages)
- Type 8792 (136 pages)
- Type 8792 (252 pages)
- Type 8718 (34 pages)
- Type 8792 (118 pages)
- Type 8791 (15 pages)
- Type 8791 (106 pages)
- Type 8791 (184 pages)
- Type 8791 (28 pages)
- Type 8791 (21 pages)
- Type 8791 (154 pages)
- Type 0911 (84 pages)
- Type 0911 (76 pages)
- Type 0911 (46 pages)
- Type 0911 (64 pages)
- Type 1058 (31 pages)
- Type 1060 (4 pages)
- Type 1066 (112 pages)
- Type 1067 (158 pages)
- Type 1077-2 (33 pages)
- Type 1094 (126 pages)
- Type 1094 (12 pages)
- Type 1094 (41 pages)
- Type 1094 (82 pages)
- Type 1115 (25 pages)
- Type 1150 (99 pages)
- Type 1541 (2 pages)
- Type 5142 (6 pages)
- Type 8619 (40 pages)
- Type 8620 (177 pages)
- Type 8622 (4 pages)
- Type 8623 (90 pages)
- Type 8623 (130 pages)
- Type 8625 (118 pages)
- Type 8624 (124 pages)
- Type 8718 (154 pages)