beautypg.com

Choosing the units for the totalizers, Configuring the analogue inputs, Refer to chap.8.9 to access parameters menu – Burkert Type 8619 User Manual

Page 73: If "mode" = "current" if "mode" = "voltage, English

background image

71

Adjustmentandcommissioning

8.10.19. choosing the units for the totalizers

The function is available if the software option "FLOW" is activated. See chap. 8.10.4.

Refer to chap. 8.9 to access Parameters menu.

This is
when the
device is be-
ing parame-
tered............
....................

Parameters

This is
when the
device is be-
ing parame-
tered............
....................

M0:Inputs

DI1/DI2

Totalizer A unit
Totalizer B unit

Mx:Inputs

The function makes it possible to select the units of the totalizers.

8.10.20. configuring the analogue inputs

Refer to chap.8.9 to access Parameters menu.

AI1/AI2

Mode:

Current

None

Voltage

Range:

0-20mA

Format:

0

20 mA:

ENTERING

Unit:

Unit list
Manual entry

This is
when the
device is be-
ing parame-
tered............
....................

Mx:Inputs

0-5V
0-10V

4-20mA

if "Mode" = "Current"

if "Mode" = "Voltage"

0.00
0.000

0.0

0 mA / 4 mA:

ENTERING

Filter:

Fast

None

Slow

5V / 10V:

ENTERING

0V:

ENTERING

if "Mode" = "Current"

if "Mode" = "Voltage"

This is
when the
device is be-
ing parame-
tered............
....................

Parameters

MODE: Choose the input signal type, current or voltage.
RANGE: Choose the range of the input signal, depending on the choice made in "MODE".
FORMAT: Choose the number of displayed decimals.
UNIT: Choose the units of the process input associated to the input signal within a predefined list or enter it.
0/4MA: Enter the value of the previously selected process variable, which is associated to a 0/4 mA input current.
Instead of being entered, the value can be automatically determined using the function "PV calibration" in the
menu "Calibration -> Mx:Inputs -> AI1 or AI2". See chap. 8.11.4.
20MA: Enter the value of the previously selected process variable, which is associated to a 20 mA input current.
Instead of being entered, the value can be automatically determined using the function "PV calibration" in the
menu "Calibration -> Mx:Inputs -> AI1 or AI2". See chap. 8.11.4.

Type 8619

english