Modifying the calibration menu access code, Adjusting the current outputs – Burkert Type 8619 User Manual
Page 86
84
Adjustmentandcommissioning
To deactivate the HOLD mode:
→
access the "HOLD" function
→
choose "Disable"
→
confirm by "OK".
8.11.2. modifying the calibration menu access code
Refer to chap. 8.9 to access Calibration menu. If the default access code "0000" is kept, the device does not
request it to access the Calibration menu.
Enter the new
CALIBRATION menu
access code
Confirm the new
code
Calibration
System
Code
0***
Confirm code
0***
8.11.3. Adjusting the current outputs
Ensure that Hold mode is deactivated before adjusting the current outputs: the
icon is displayed.
Refer to chap. 8.9 to access Calibration menu.
Calibration
M0:Outputs
AO1/AO2
4mA
20mA
ENTERING
ENTERING
Mx:Outputs
4mA: Adjust the offset of the current output.
When the "4mA" function is selected, the multiCELL generates a current of 4 mA:
→
measure the current emitted by the 4-20 mA output using a multimeter;
→
enter the value given by the multimeter.
20mA: Adjust the span on current output 1 or current output 2.
When the "20mA" function is selected, the multiCELL generates a current of 20 mA:
→
measure the current emitted by the 4-20 mA output using a multimeter;
→
enter the value given by the multimeter.
→
Type 8619
english