43
Adjustmentandcommissioning
TITLE: Entering the name displayed in the corresponding "Ux" view. See chap. 8.4.
1
13:40
3
0
1
7 pH
1 S/cm
205 l/min
43 °C
U3:PROCESS1
The title of the view is displayed
here.
MENU
Indicates the number of
the board or module from
which the process value
originates.
2010/06/29
Figure 28 :
Example of a title for a customised view
LINE1 to LINE4: Setting the parameters for the values (1, 2 or 4) displayed in a customised "Ux" view:
-
PV: Select the digital input, the analogue output or the physical parameter to be displayed on the line selected in this
customised view. The options available depend on the modules fitted.
One of the "PVs" in the conductivity module, available for the customised "Ux" views, is "USP" (see
chap. 8.10.24).
-
UNIT: Select the unit in which the digital input, the analogue output or the physical parameter selected is displayed on
the PV function above.
-
FILTER: Select the level of damping for the measurement signal on the digital input, the analogue output or the physical
parameter displayed on the line selected. Three levels of damping are proposed: "slow" (slow filtering has a high damp-
ing effect), "fast" or "none"
"Slow" filter
"Fast" filter
"No" filter
150 ms
6 s
t
30 s
t
t
Figure 29 :
Types of filters and associated damping effects
1
13:40
U1:PH
7 pH
MENU
2010/06/29
2
3
2
3
13:40
7 pH
1 S/cm
U2:PH_COND
MENU
2010/06/29
1
13:40
3
0
1
7 pH
1 S/cm
205 l/min
43 °C
U3:PROCESS1
MENU
2010/06/29
Figure 30 :
Examples of customised views with 1, 2 and 4 lines
Type 8619