beautypg.com

Burkert Type 8619 User Manual

Page 92

background image

90

Adjustmentandcommissioning

Detailed procedure for teach-in by flow rate

Choose the flow rate unit in which teach-in is run:

Mx:Inputs

Calibration

DI1/2: Flow

M0:Inputs

Flow unit

Flow teaching

L/h

Igal/s

...

Run teach-in by flow rate:

Allow the fluid to circulate in the pipes and wait
for the flow rate to stabilise.

The multiCELL displays the K factor
calculated by teach-in.

OK

OK

OK

The multiCELL displays the flow rate
calculated with the current K factor

Enter the value of the flow rate
in the pipe.

0049 L/min
0001 L/s

Flow teaching

Calibration

DI1:/DI2:Flow

M0:Inputs

Start teaching

Meas. in progr.

Average flow

Flow K factor:
K=3,810

Teaching result

0005 L/min

The multiCELL calculates the flow rate
measured in the pipe, with the current K factor,
for 100 seconds

1)

.

The bargraph shows the time passed.

1)

Measurement can be suspended at any time by clicking on "OK".

Type 8619

english