Contact low high process input ("pv") not inverted, Contact low high process input ("pv") inverted, English – Burkert Type 8619 User Manual
Page 77
75
Adjustmentandcommissioning
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
M0:Outputs
DO1/DO2
Mode:
Parameters
Mx:Outputs
PFM
1)
M1:
M0:MAIN
...
Mx:
PV:
Mode:PFM
Pulse width:
Invert:
No
Yes
ENTERING
Low:
ENTERING
High:
ENTERING
Max. freq.:
ENTERING
Pulse
DI2
DI1
Input:
Mode:Pulse
Pulse:
ENTERING
2)
1)
The choices offered depend on the modules fitted and/or the options activated. See chap. 8.10.4 and chap. 8.16.
2)
Function possible only for the DO1 and DO2 outputs on the M0:MAIN board
MODE: Choose the switching operating for the digital output selected.
configuring in "on/off" operating
PV: Choose a process input with 2 states, ON or OFF, associated with the output.
One of the "PVs" on the M0:MAIN board, available in "On/Off" operating, is the "warning" event. Such
an event is generated when the calibration date expires and/or during the monitoring of certain process
parameters (see chap. 8.11.8, 8.11.9. and 8.12.4 to 8.12.6). When the "warning" event is associated with
a digital output, the output switches whenever such an event is generated by the multiCELL. See also
chap. 9.3
One of the "PVs" in the conductivity module, available in "On/Off" operating, is "USP" (see
chap. 8.10.24).
INVERT: Invert the output or not.
DELAY: Choose the value of the time-out before the output is switched.
configuring in "hysteresis" operating
The output status changes when a threshold is reached:
• by increasing process input value, the output status changes when the high threshold is reached.
• by decreasing process input value, the output status changes when the low threshold is reached.
ON
OFF
contact
Low High process input ("PV")
not inverted
ON
OFF
contact
Low High process input ("PV")
inverted
Figure 63 :
Hysteresis operating
PV: Choose the process input associated with the output.
LOW: Choose the value of the low switching threshold of the output.
Type 8619
english