Specifications, Assessing if the lpc is free of particles – Ocean Optics S2000 User Manual

Page 71

background image

Assessing if the LPC is Free of Particles

Fluids need to be relatively particle-free. Particles larger than 20

µ

m can be trapped inside the tubing and can then

block or scatter a significant amount of light. To rid the LPC of particles, follow these steps:
1.

Pump the sample fluid through the LPC.

2.

While in Scope Mode, save a dark spectrum with the light source off and a reference spectrum with the light
source on.

3.

Continue to pump the sample fluid and switch to the Absorbance Mode. Ideally, you should see a spectrally
flat line. Particle effects manifest themselves as an exponentially decreasing curve from shorter to longer
wavelengths. The length of time that you pump the sample and the magnitude of the absorbance peak, depends
upon the time required and the minimum detectable absorbance value for your specific analysis. Pre-filtering
of the sample may be required to eliminate this exponentially decreasing absorbance spectrum if it is
significant to your analysis.

Specifications

Sampling Chambers: LPC

- 71 -

Path lengths:

1-meter, 5-meter and 10-meter options (standard); custom
lengths also available

Tubing:

Teflon

Amorphous Fluoropolymer 2400

~560

µ

m inner diameter, ~800

µ

m outer diameter

Refractive index:

1.29

Internal volume:

250

µ

l/meter

Chemical resistance:

tubing can be altered by perfluorinated solvents, FREON

113,

and Perclene

Recommended optical fibers for coupling to
spectrometers, light sources:

400

µ

m illumination fiber (UV-VIS)

200

µ

m or 400

µ

m read fiber (UV-VIS)

Plumbing fittings:

standard ¼" x 28 chromatography fittings