Specifications – Ocean Optics S2000 User Manual
Page 56
Specifications
*
Though the VIS-NIR lens is optimized for use to 2
µ
m, it can be configured to "see" only to 1100 nm with our
S2000 spectrometer.
**
There is no single combination of optical fibers that will satisfy the requirements of every application. As a
rule, however, it is best to use a large-diameter (>50
µ
m) illumination fiber to get the maximum light
throughput, and a small-diameter (<50
µ
m) read fiber to achieve the best optical resolution.
Sampling Chambers: CUV-UV, CUV-VIS
- 56 -
Path length:
1 cm
Collimating lenses (VIS-NIR):
BK 7 glass (~360 nm - 2
µ
m*), 5 mm diameter, f/2
Collimating lenses (UV-VIS-NIR):
Dynasil 1100 quartz (200 nm-2
µ
m*), 5 mm diameter, f/2
Collimating lens termination:
SMA 905
Collimating lenses assembly (sample
compartment) dimensions:
2.0" x 1.5" (LW)
Filter slot:
accepts filters up to ¼" (6 mm) in thickness
Base material:
aluminum
Base length:
5.5"
Connections in base:
water inlet
water outlet
extra connection
Water input fittings:
1/8" NPT pipe thread
Typical optical fibers specified for optimum
performance (light throughput and optical
resolution)**:
200
µ
m illumination fiber
50
µ
m read fiber