Spectrometer Configuration Dialog Box

Now that OOIBase32 is running, you need to configure your system. Select Spectrometer | Configure from the
menu. Go through each page in the Spectrometer Configuration dialog box to set system parameters. (See the

OOIBase32 Spectrometer Operating Software Manual

for details.)

"

In the Wavelength Calibration page, the
coefficients for each spectrometer channel in
your system have already been loaded as part
of the spectrometer configuration file. Check
the Enabled box for each spectrometer
channel in your system.

"

In the A/D Interface page, enter the same
values as you did in the Configure Hardware
dialog box.

"

The Detector Linearity page in this dialog
box allows you to enter coefficients for an
algorthim that corrects for rare occurances of
non-linearity of the detector. Contact Ocean
Optics for more information.

Save the spectrometer configuration file by
choosing Spectrometer | Save Configuration As
from the OOIBase32 menu. You can rename the file or use the default file name (

[your serial number].spec

). You

will then be asked if you would like to make this file the default spectrometer configuration file. Choose Yes. The
next time you run OOIBase32, the software will use the file as the standard for your spectrometer configuration.
When you exit OOIBase32, any changes to the configuration file will be automatically saved to the default file.

OOIBase32 Settings Dialog Box

At this point, it is a good idea to configure several OOIBase32 operation parameters. Choose Edit | Settings from
the menu to open the OOIBase32 Settings dialog box. Go through each page of this dialog box to select options
for saving, opening, and printing data; to configure default setting files; and to select other important options such
as storing and copying data and choosing warning messages. (See the

OOIBase32 Spectrometer Operating

Software Manual

for details.)

Configure Data Acquisition Dialog Box

Finally, select Spectrum | Configure Data Acquisition from the menu to set your data acquisition parameters in
the Configure Data Acquisition dialog box. The Basic page allows you to set the integration time and choose
averaging and boxcar smoothing values. The External Trigger page allows you to specify the external trigger
mode. The Strobe page allows you to control external strobe events with the spectrometer. (See the

OOIBase32

Spectrometer Operating Software Manual

for details.)

Specifications

A/D Converters: ADC1000 (and PC2000)

- 22 -

Board architecture/design:

half-length ISA-bus card for single slot in desktop PC

Resolution:

12-bit

Sampling frequency:

1 MHz (maximum)

Interface cable:

For the ADC1000: 37-pin connector to A/D card and 25-pin connector to
spectrometer

Multiple-channel capability:

up to 8 spectrometer channels for S2000/PC2000 spectrometers

Installation parameters:

set dip switch for Base Address (default = 768 decimal, 300 hexadecimal)
set dip switch for IRQ (default = 7)

Integration time:

3 milliseconds to 60 seconds with S2000 series spectrometers