R-ls-450 operating matrix, Jumper block 3 (jp3) – Ocean Optics S2000 User Manual
Page 47
Jumper Block 3 (JP3)
There are two sets of pins in JP3. The jumper position here determines the source of control for the R-LS-450:
manual or remote control. A jumper over the Remote pins means that you can control the R-LS-450 through the
software (if other jumper blocks are configured correctly).
R-LS-450 Operating Matrix
This matrix will help you configure the jumper blocks on the R-LS-450.
Using JP3 on the S2000
You can also have functionality of the R-LS-450 by utilizing Jumper Block 3 (JP3) on the S2000 circuit board. The
S2000’s JP3 allows you to control the R-LS-450 through the OOIBase32 software.
Continuous Wave Mode with the S2000’s JP3
You can control the turning on and off of the R-LS-450 while it is in continuous wave mode through the Strobe
Enable function in the OOIBase32 software if
"
The switch is turned to continuous wave mode
"
There is a jumper over the pins in JP1 of the R-LS-450
"
There is a jumper over the CW pins in JP2 of the R-LS-450
"
There is a jumper over the Remote pins in JP3 of the R-LS-450
"
There is a jumper over the 2 pins in JP3 of the S2000 board
Pulsed Mode with the S2000’s JP3
Another way you can control the pulses per second of the R-LS-450 is if
"
The switch is turned to pulsed mode
"
There is a jumper over the pins in JP1 of the R-LS-450
"
There is a jumper over the CS pins in JP2 of the R-LS-450
"
There is a jumper over the over the Remote pins in JP3 of the R-LS-450
"
There is a jumper over pins labeled /16, /14, /12, or /10 in JP3 of the S2000, depending on the pulse rate you
need. The pulses per second are also dependent upon the frequency of your A/D card. See the table below
for choices.
Light Sources: R-LS-450
- 47 -
S1 Switch
JP1
JP3
LED Status
Off
No jumper
No jumper
Off
CW
No jumper
No jumper
Continuously on
CW
Jumpered
Jumper Remote pins
Continuous wave mode controlled by software
(see
Continuous Wave Mode with the S2000’s JP3 for more
information)
CW
Jumpered
Jumper Manual pins
Continuously on
Pulsed
No jumper
No jumper
Pulse rate determined by JP2 on the R-LS-450 board
(see the JP2 table for pulse rates)
Pulsed
Jumpered
Jumper Remote pins
Pulsed mode controlled by software
(see
Pulsed Mode with the S2000’s JP3 for more information)
Pulsed
Jumpered
Jumper Manual pins
Pulse rate determined by JP2 on the R-LS-450 board
(see the JP2 table for pulse rates)
Pins on the S2000’s JP3
DAQ700
ADC500, SAD500
ADC1000
/16
1.5
7.6
15.2
/14
6.1
30.4
60.8
/12
24.0
122.0
244.0
/10
98.0
488.0
976.0