S2000 specifications – Ocean Optics S2000 User Manual
Page 12
S2000 Specifications
S2000 Miniature Fiber Optic Spectrometers
- 12 -
Absolute Maximum Ratings
V
CC
+ 5.5 VDC
Voltage on any pin
V
CC
+ 0.2 VDC
Physical Specifications
Physical dimensions (no enclosure)
5.47” x 3.90” x 0.75” LWH (master only)
139 mm x 99 mm x 19 mm LWH (master only)
Physical dimension (with enclosure)
5.63” x 4.09” x 1.58” LWH (master only)
143 mm x 104 mm x 40 mm LWH (master only)
Weight
200 g (master only, no enclosure)
Power
Power requirement (master)
130 mA at +5 VDC
Power requirement (slave)
70 mA at +5 VDC
Supply voltage
4.5 – 5.5 V
Power-up time
3 msec
Spectrometer
Design
asymmetric crossed Czerny-Turner
Focal length (input)
42 mm
Focal length (output)
68 mm (75, 83 and 90 mm focal lengths are also available for some
configurations)
Input fiber connector
SMA 905
Gratings
14 different gratings
Entrance slit
5, 10, 25, 50, 100, or 200
µ
m slits. (Slits are optional. In the absence
of a slit, the fiber acts as the entrance slit.)
Detector
Sony ILX511 CCD
Filters
2
nd
order rejection, long pass (optional)
Spectroscopic
Integration time
3 – 65,000 msec
Dynamic range
2 x 10
8
Signal-to-Noise
250:1 single acquisition
Readout noise (single dark spectrum)
3.5 counts RMS, 20 counts peak-to-peak
Resolution (FWHM)
0.3 – 10.0 nm varies by configuration
Stray light
<0.05% at 600 nm; <0.10% at 435 nm
Spectrometer channels
8 (master plus up to 7 slaves)
Environmental Conditions
Temperature
-30
°
- +70
°
C Storage; -10
°
- +60
°
C Operation