beautypg.com

Specifications – Ocean Optics S2000 User Manual

Page 65

background image

6.

Gently separate the two halves of the lamp and pull the bulb out of its housing.

7.

Detach the wire and socket from the lamp leads. Remove bulb unit and discard.

8.

Plug the new bulb into the socket and slide it forward into the front of the lamp as far as it will go.

9.

Tighten the set screw on the bottom of the lamp to hold the bulb in place.

10. Close the two halves of the lamp, being careful not to pinch the wires. Replace the two screws at the back of

the lamp and secure the fan to the bottom of the lamp.

Specifications

Sampling Chambers: ISS-2

- 65 -

Path length:

1 cm

Collimating lens:

BK 7 glass (~360 nm - 2

µ

m*), 5 mm diameter, f/2

Collimating lens termination:

SMA 905

Filter slot:

accepts filters up to ¼" (6 mm) in thickness

Base material:

aluminum

Spectral range:

360 nm - 2

µ

m*

Dimensions:

9.0 cm x 5.0 cm x 3.2 cm (LWH)
3.5" x 2.0" x 1.25" (LWH)

Power input:

12 VDC/800 mA, 7-20 VDC/0.5-2 amps

Power output:

6.5 watts

Bulb life:

900 hours

Bulb color temperature:

3100K

Output to bulb:

5 volts/1.3 amps

Output regulation:

0.2% voltage

Time to stabilized output:

~ 30 minutes

Bulb output:

7400 foot-candles (7.4MSCP)

*

Though the product can be used to 2

µ

m, it can be configured to "see" only to 1100 nm with our S2000

spectrometer.