Setting the integration time – Ocean Optics S2000 User Manual

Page 48

background image

You can control the pulses per second of the R-LS-450 through the Flash Delay function in the OOIBase32
software if:

"

The switch is turned to pulsed mode

"

There is a jumper over the pins in JP1 on the R-LS-450 board

"

There is a jumper over the pins labeled CS in JP2 on the R-LS-450 board

"

There is a jumper over the pins labeled Remote in JP3 on the R-LS-450 board

"

There is a jumper over the pins labeled 2 in JP3 on the S2000 board

Setting the Integration Time

When using any of the pulsed modes for the R-LS-450, you need to ensure that a constant number of flashes occur
for every integration cycle. This achieves a continuous and stable signal. The integration time is set in the
OOIBase32 software. To achieve a constant number of flashes per integration cycle, the integration time must be a
multiple of those shown in the following table, according to the A/D converter being used:

Light Sources: R-LS-450

- 48 -

Pins on the JP3

(of the S2000)

Integration time for

DAQ700 must be

multiple of

Integration time for

ADC500 / SAD500 must be

multiple of

Integration time for

ADC1000 must be

multiple of

/16

512

128

64

/14

128

32

16

/12

32

8

4

/10

8

(with a min. value of 24 ms)

4

4

2

N/A

N/A

N/A