Installing Filters

1.

Loosen the filter clamping screw with the provided screwdriver.

2.

Insert the filter into the filter slot. The filter slot can accommodate filters up to 6 mm thick.

3.

Clamp the filter in place by gently tightening the clamping screw finger tight.

Using the Temperature Stabilization Feature

This feature is used to heat or cool the cuvette holder base and cuvette.
1.

Remove the two plugs from the top side of the base. (The plug on the right side of the base should stay in
place but might require thread tape.)

2.

Replace the plugs with the two barbed fittings (or any 1/8” NPT pipe thread fittings). Thread tape might be
required on the fittings.

3.

Connect the fittings to a water source. Water will circulate through the base.

Specifications

*

Though the VIS-NIR lens is optimized for use to 2

µ

m, it can be configured to "see" only to 1100 nm with our

S2000 spectrometer.

Sampling Chambers: CUV-ALL

- 59 -

Path length:

1 cm

Collimating lenses (VIS-NIR):

(2) BK 7 glass (~360 nm - 2

µ

m*), 5 mm diameter, f/2

Collimating lenses (UV-VIS-NIR):

(2) Dynasil 1100 quartz (200 nm - 2

µ

m*), 5 mm diameter, f/2

Collimating lens termination:

SMA 905

Fluorescence option:

position 2 collimators at 90

°

Filter slot:

accepts filters up to ¼" (6 mm) in thickness

Base material:

aluminum

Water input fittings:

1/8" NPT pipe thread