Appendix a specifications, Appendix a, Specifications – National Instruments AT-MIO-16X User Manual
Page 224: Analog input
© National Instruments Corporation
A-1
AT-MIO-16X User Manual
Appendix
A
Specifications
This appendix lists the specifications of the AT-MIO-16X. These are
typical at 25° C unless otherwise stated. The operating temperature
range is 0° to 70° C. A warmup time of at least 15 minutes is required.
Analog Input
Number of input channels ...................16 single-ended, 8 differential
Analog resolution ...............................16-bit, 1 in 65,536
Maximum sampling rate .....................100 ksamples/sec
Relative accuracy ................................±1.5 LSB maximum over
temperature,
±0.75 LSB typical
Differential nonlinearity (DNL) ..........±1 LSB maximum (no missing
codes over temperature,
±0.5 LSB typical
Differential analog input ranges ..........±10 V or 0 to +10 V,
software-selectable
Common-mode input range .................Each input to the
instrumentation amplifier
should remain within ±11 V of
AIGND at any gain or range
Overvoltage protection
(ACH0 to ACH15 and AISENSE).......±15 V power off,
±25 V power on
Common-mode rejection ratio.............105 dB, all gains DC to 60 Hz
Bandwidth (–3 dB)..............................DC to 255 kHz, all gains
Input bias current ................................±1 nA