Chapter 5

Programming

AT-MIO-16X User Manual

5-34

© National Instruments Corporation

Programming the Waveform Cycle Counter

Select the appropriate counter (1, 2, or 5) from the Am9513A
Counter/Timer to be used for counting DAC buffer cycles. To program
the cycle counter, complete the following programming sequence. All
writes are 16-bit operations. All values given are hexadecimal.

1.

Write FF00 + n to the Am9513A Command Register to select the
Counter n Mode Register.

2.

Write 0325 to the Am9513A Data Register to store the Counter n
mode value. Am9513A counter mode information can be found in
Appendix C, AMD Am9513A Data Sheet.

3.

Write (FF08 + n) to the Am9513A Command Register to select the
Counter n Load Register.

4.

Write the desired cycle count to the Am9513A Data Register to
store the Counter n load value.

5.

Write the following value to the Am9513A Command Register to
load Counter n.

FF41 — Load Counter 1

FF42 — Load Counter 2

FF50 — Load Counter 5

6.

Write (FFF0 + n) to the Am9513A Command Register to decrement
Counter n.

7.

Write the following value to the Am9513A Command Register to
arm Counter n.

FF21 — Arm Counter 1

FF22 — Arm Counter 2

FF30 — Arm Counter 5

After you complete this programming sequence, Counter n is
configured to count the DAC buffer retransmit signal from SOURCE3
as soon as the load/arm counter command is written.