© National Instruments Corporation

It is important to realize that inaccuracy of the internal voltage
reference results only in gain error. Offset error is unaffected. If an
application can tolerate slight gain inaccuracy, there should not be a
need to redetermine the value of the onboard reference.

Calibration DACs

There are eight 8-bit DACs (CALDAC<0..7>) and one 12-bit DAC
(CALDAC8) on the AT-MIO-16X that are used for calibration. These
DACs are described in Table 6-2.

Reference Calibration

The AT-MIO-16X has a stable voltage reference to which gain can be
calibrated. The value of this voltage reference is determined through the
reference calibration routine, which requires a known external voltage
between 5 and 9.99 V to be connected differentially on any desired
input channel. The routine calibrates the circuitry to the external
reference and then reads the internal reference. This value is stored as

Table 6-2. Calibration DACs

Analog

DAC

Function

Adjustment

Range

Incremental

Effect

Input

CALDAC0

Coarse gain trim

±1169 ppm

–9.13 ppm

CALDAC1

Fine gain trim

±65.1 ppm

–0.509 ppm

CALDAC2

Coarse postgain offset trim

±51.8 ppm

404

µ

V

CALDAC3

Fine postgain offset trim

±512

µ

V

4.00

µ

V

CALDAC8

Pregain offset trim

±1.2

µ

V

0.586

µ

V

Output

CALDAC4

DAC0 gain trim

±976 ppm

7.62 ppm

CALDAC5

DAC1 gain trim

±976 ppm

7.62 ppm

CALDAC6

DAC0 offset trim

±4.86 mV

38.0

µ

V

CALDAC7

DAC1 offset trim

±4.86 mV

38.0

µ

V