Table 4-10. analog output voltage versus digital c, Table 4-11. analog output voltage versus digital c, Table 4-10 – National Instruments AT-MIO-16X User Manual
Page 138: Table 4-11
Chapter 4
Register Map and Descriptions
AT-MIO-16X User Manual
4-42
© National Instruments Corporation
The formula for the voltage output versus digital code for a bipolar
analog output configuration in two’s complement form is as follows:
V
out
= V
ref
* (digital code)
32,768
where V
ref
is the positive reference voltage applied to the analog output
channel. The digital code in the preceding formula is a decimal value
ranging from –32,768 to +32,767.
Table 4-10. Analog Output Voltage Versus Digital Code (Unipolar Mode)
Digital Code
Voltage Output
Decimal
Hex
V
ref
= 10 V
0
0000
0.0 V
1
0001
152.6
µ
V
16,384
4000
2.5 V
32,768
8000
5.0 V
49,152
C000
7.5 V
65,535
FFFF
9.999847 V
Table 4-11. Analog Output Voltage Versus Digital Code (Bipolar Mode)
Digital Code
Voltage Output
Decimal
Hex
Reference = 10 V
–32,768
8000
–10.0 V
–32,767
8001
–9.999695 V
–16,384
C000
–5.0 V
–1
FFFF
–305.2
µ
V
0
0000
0.0 V
1
0001
305.2
µ
V