3-49

displayed value for slew key increment will only show two decimal places, however the
true four-decimal place value is stored internally and is the actual value added/sub-
tracted when a slew key is pressed. (For example, entering an increment of “.0055”

will result in a displayed value of “0.00;” pressing the slew key continuously for a

period of time in “Output is OFF” or Keypad control” status will verify that the pro-
grammed or actual value is changing.)

4.

Press ENTER to accept the new value; the display shows the new slew key increment val-
ues.

NOTE: If a value greater than 1 (V or A) is entered, the adjustment will be set to 1 (maximum

value) without any warning message. Repeat step 3. to readjust, as desired.

5.

Press MENU ON/OFF to return to the menu level. Press it again to exit the menu system.

3.2.2.7

REMOTE DISPLAY BLANKING

The front panel display may be blanked or disabled during remote operation. This may improve
the response time of the MBT unit during quick, successive remote operations. The dispaly may
be disabled either temporarily for a selected number of seconds, or continuously once a remote
command is received, until the unit is powered off. While the display is disabled, it will show the
message "REMOTE CONTROL".

1.

Press MENU ON/OFF to enter the menu, then NEXT until "Rem Dsp Blanking" is dis-
played; press ENTER to select this function.

2.

The unit will display "Don't Blank," "Stay Blank," or the number of seconds (1, 3,

or 10) to keep the display disabled. Use the slew keys to select the mode that you prefer.

3.

Press MENU ON/OFF to return to menu level. Press it again to exit menu system.

3.2.2.8

GPIB ADDRESSING MODE

The MBT-G supports either Primary or Secondary addressing mode for GPIB remote program-
ming. In the Secondary addressing mode, the secondary address is the channel number to
which the command will be sent. Otherwise, the channel number is selected by use of the
"INSTR" command, or by appending the channel number to the command itself.

1.

Press MENU ON/OFF to enter the menu, then NEXT until "GPIB addressing" is dis-

played; press ENTER to select this function.

2.

The display shows either "Pri.Addressing" or "Sec.Addressing." Use the slew

keys to toggle between the two modes.

3.

Press MENU ON/OFF to return to menu level. Press it again to exit menu system.

3.2.2.9

COMPATIBLITY MODE

The MBT-G has 2 compatibilty modes, Off and ON. When selected to be ON, the unit is back-
ward compatible with older Kepco products. When OFF, the unit is completely compatible with