PRELIMINARY OPERATIONAL CHECK

A simple operational check after unpacking and before equipment installation is advisable to
ascertain whether the power supply has suffered damage resulting from shipping. Refer to Fig-
ures 2-1 and 2-3 for location of operating controls and electrical connections. Tables 3-1 and 3-
2 ex
plain the functions of operating controls/indicators and keypad keys, respectively.

WARNING

BEFORE APPLYING THE A-C SOURCE TO THE POWER SUPPLY, VER-

IFY THAT THE LINE VOLTAGE TO BE SUPPLIED MATCHES THE POSI-

TION OF THE AC INPUT SELECTOR SWITCH AT THE REAR PANEL

(FACTORY DEFAULT IS 115V).

1. With no load connected, set POWER circuit breaker to the ON position. As shown below, the

alphanumeric display indicates the model and options; VOLTAGE MODE status indicator is
on, all other status indicators are off.

2. After approximately 2 seconds, the display changes to:

3. Rotate VOLTAGE control first clockwise, then counterclockwise; verify voltage shown on dis-

play increases for clockwise rotation and decreases for counterclockwise rotation. If the volt-
age control has no effect, refer to PAR.s 3.2.2 and 3.2.2.2 to check whether the adjustments
are locked; unlock adjustments if necessary.

NOTE: If the voltage (or current) does not increase beyond a certain point, press SHIFT then

.” keys (STATUS command) to see the overvoltage protection (OVP) and overcurrent
protection (OCP) values. Refer to PAR.s 3.2.2 and 3.2.2.5 set new values if necessary.

4. Verify short beeps are audible when VOLTAGE control is rotated. If beeps are not heard

refer to PAR.s 3.2.2 and 3.2.2.1 to check whether the speaker is disabled; enable speaker if
necessary.

5. Rotate CURRENT control first clockwise, then counterclockwise; verify current shown on

display increases for clockwise rotation and decreases for counterclockwise rotation. Verify
short beeps are audible when CURRENT control is rotated.

VOLTAGE

MODE

CURRENT

MODE

POLARITY

REVERSED

REMOTE

OUTPUT

ENABLED

STATUS

KEPCO, INC.

MBT 100-3.6MGR

VOLTAGE

MODE

CURRENT

MODE

POLARITY

REVERSED

REMOTE

OUTPUT

ENABLED

STATUS

Output is OFF

***V ***A

NOTE: ***V and ***A are the output volt-

age and current settings in effect
at time of last operation (for first
time use, the voltage and current
are set to 0).