beautypg.com

Table 3-17. sample program (assigned program #2), 17 sample program (assigned program #2) -35, Stp 3 end of program pr#2 stp: 0 sc:0 – KEPCO MBT Series User Manual

Page 73: Const volt, 0v 0.00a

background image

MBTSVC111609

3-35

The following step-by-step procedure explains how to enter a simple program to illustrate the
programming technique. The program to be entered is shown in Table 3-17; the values are arbi-
trary, for instructional purposes only.

1.

Press SHIFT, then press (ADD STEP) to add step 0. The display shows one of the
screen listed in Table 3-15 (depending upon the last screen displayed during a prior pro-
gramming session).

NOTE: The screens may be programmed in any order. It is not necessary to change values if

the default values are correct, however it is recommended that all screens be checked.

2.

Repeat step 1 above twice to add steps 1 and 2 to the program.

3.

Press NEXT

4.

Press NEXT again to display step 0.

5.

You are ready to begin changing the default values of the added steps to new values.
Press Vadj

TABLE 3-17. SAMPLE PROGRAM (ASSIGNED PROGRAM #2)

PROGRAM

STEP

(Stp: )

SCREEN 0

SCREEN 1

SCREEN 2 *

SCREEN 3 *

SCREEN 4 *

VOLTAGE

CURRENT

DELAY

MODE *

POLARITY *

OUTPUT *

0

1.0V

2.1A

20s

VOLTAGE

NON-

REVERSED

ENABLED

1

2.0V

1.3A

15s

VOLTAGE

REVERSED

ENABLED

2

5.2V

3.3A

10s

CURRENT

NON-

REVERSED

DISABLED

*

MR and MGR Models only.

NOTE:

It is recommended that programs be entered either by first adding the “blank” steps, then programming the values for
each screen of each step, or alternatively, by adding step 0, programming the values for each screen of step 0, then
adding step 1, programming the values for step 1, etc. There are, however, no restrictions on the sequence of program-
ming steps other than “blank” steps must be added before they can be assigned values.

Stp 3
End of program

Pr#2 Stp: 0 Sc:0

***V ***A

NOTE: If Sc: 0 is not displayed, use or to

cycle through the screens until Sc:0 is
displayed. (Remember, even though we
are programming screen 0 first, the
screens and steps may be programmed
in any order.)

Const VOLT =

0.0V 0.00A